Cleaning, Troubleshooting, Service information – Ohaus MB23 MOISTURE ANALYZER Manual it User Manual
Page 9: Accessories

MB23 / MB25 EN-7
Cleaning
Disconnect the Analyzer from the power supply before cleaning the Analyzer.
Make sure that no liquid enters the interior of the Analyzer.
Make sure the Analyzer is cooled down before cleaning.
Clean the Analyzer at regular intervals.
Housing surfaces and the temperature sensor may be
cleaned with a lint-free cloth slightly dampened with water
or a mild cleaning agent.
Glass surfaces may be cleaned with a commercial glass
cleaner.
Do not use solvents, harsh chemicals, ammonia or
abrasive cleaning agents.
Troubleshooting
Symptom / Display
Possible Cause
Remedy
Cannot turn on
No power to analyzer
Verify connections, fuse and voltage
Weight value blinking Sample weight less than 0.5g
Increase sample size
“Tare” shown
Pan weight needs set to zero
Press Tare
“Close Cover” shown
Cover needs to be closed before starting test
Close cover
Poor accuracy
Improper calibration
Unstable environment
Perform calibration
Move the analyzer to suitable location
Cannot calibrate
Unstable environment
Incorrect calibration masses
Move the analyzer to suitable location
Use correct calibration masses
Err 7.0
Time out
Err 8.1
Pan support has load during power on
Remove weight from pan support
Err 8.2
Pan support was removed prior to power on
Install pan support
Err 8.3
Weight on pan support exceeds capacity
Remove weight from the pan support
Err 8.4
Pan support was removed during weighing
Re-install pan support
Err 8.5
Cannot perform temperature calibration
Cool down the analyzer
Err 8.6
Current temperature is higher than setting
Cool down the analyzer
Err 9.5
Factory calibration data corrupted
Contact an Ohaus-authorized dealer
Err 10.3
Temperature sensor out of range – high
Contact an Ohaus-authorized dealer
Err 10.4
Temperature sensor out of range – low
Contact an Ohaus-authorized dealer
Err 10.5
Temperature sensor defective
Contact an Ohaus-authorized dealer
Err 10.6
Temperature sensor defective
Contact an Ohaus-authorized dealer
Err 10.7
Temperature out of control
Contact an Ohaus-authorized dealer
Err 11.0
AC power is unstable
Use a power line conditioner
Err 53
EEPROM checksum error
Contact an Ohaus-authorized dealer
Err 54
Loading of calibration parameters failed
Contact an Ohaus-authorized dealer
Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized OHAUS service agent.
Please visit our web site: www.ohaus.com to locate the OHAUS office nearest you.
Accessories
Description
Part No.
Security Locking Cable
76288-01
Security Lock (Kensington
®
type)
470004-010
Temperature Calibration Kit
11113857
Sample Pans (50/Box)
80850086
Pads, Glass Fiber (200/Box)
80850087
Reusable Sample Pans (3/pk)
80850088
Reusable Deep Sample Pans (3/pk)
80252479
Reusable Wire Cage
80252477
Sample Pan Handler
80252476
50g Calibration Weight
51054-16
Description
Part No.
STP103 Thermal Printer, US
80251992
STP103 Thermal Printer, EU
80251993
STP103 Thermal Printer, UK
80251994
CBM910 Impact Printer, JP
80252041
CBM910 Impact Printer, US
80252042
CBM910 Impact Printer, EU
80252043
Cable, STP103
80252581
Cable, CBM910
80252571
Data collection Software
SW12
Protective glass
Temperature sensor