beautypg.com

Calibration – Ohaus DS Series Electronic Digital Bench Scales User Manual

Page 32

background image

9

Calibration

Calibration

Calibration

Calibration

Calibration

NOTE

NOTE

NOTE

NOTE

NOTE: This procedure requires that
the scale must be ON

must be ON

must be ON

must be ON

must be ON. Calibration can

be in made in pounds or kilograms.

1.

Remove all weights from the
platform.

2.

Press

ON/OFF/RE-ZERO

to

turn ON the scale. The display
should show either 0.00 kg or
0.00 lb.

3.

Press

ON/OFF/RE-ZERO

and

hold for five seconds

five seconds

five seconds

five seconds

five seconds until CO

appears in the display. Wait for
display C100.

4.

Place a test weight or weights
on the platform which equals
the full capacity of the scale.
For example, on Model DS4, a
weight of 20 kg would be used.
Press

ON/OFF/RE-ZERO

and

wait for display 100.

5.

The scale is readjusted and
ready for operation after press-
ing the

ON/OFF/RE-ZERO

button once again.