Calibration – Ohaus DS Series Electronic Digital Bench Scales User Manual
Page 17

9
Calibration
Calibration
Calibration
Calibration
Calibration
NOTE
NOTE
NOTE
NOTE
NOTE: This procedure requires that
the scale must be ON
must be ON
must be ON
must be ON
must be ON. Calibration can
be in made in pounds or kilograms.
1.
Remove all weights from the
platform.
2.
Press
ON/OFF/RE-ZERO
to
turn ON the scale. The display
should show either 0.00 kg or
0.00 lb.
3.
Press
ON/OFF/RE-ZERO
and
hold for fifteen seconds
fifteen seconds
fifteen seconds
fifteen seconds
fifteen seconds, the
display will show 0.00 kg or
0.00 lb. Release the
ON/OFF/
RE-ZERO
button, the display
will show in sequence 0 then
000 followed by CAL.
NOTE
NOTE
NOTE
NOTE
NOTE: If more than one weight is
used to calibrate, place them
on the scale at the same time.
4.
Place a weight or weights on
the platform which equals the
full capacity of the scale. For
example, on Model DS4, a
weight of 20 kg would be used.
When the display indicates the
weight on the platform, remove
the weight(s).
5.
Momentarily press the
ON/
OFF/RE-ZERO
button, the
scale should now be calibrated
and the display should indicate
0.00kg
NOTE
NOTE
NOTE
NOTE
NOTE: The display will alternate
between 000
000
000
000
000 and CAL
CAL
CAL
CAL
CAL all the
time until a weight is placed on
the platform or until the scale
turns off by the auto-shut-off
feature.