Activeforever Mettler Sonicator Plus 992 2-Channel Combination Therapy Device User Manual
Page 2

2
Treatment timer:
Timer Accuracy:
±0.5 minutes for times less than 5 minutes
±10% for times from 5 to 10 minutes
±1.0 minute for times greater that 10 minutes
±5%, CE specification
Maximum Treatment Time:
60 minutes–electrical stimulation
30 minutes–ultrasound or combination therapy
Treatment Timer:
Treatment time counts down to zero when a time is set, or up
to 60 or 30 minutes when no time is set. The digital timer
indicates time in minutes and seconds. The timer also
indicates the remaining or elapsed treatment time during the
“Hold” period.
Ultrasonic Generator Specifications:
Frequency:
1.0 MHz ±5%
3.2 MHz ±5%
3.3 MHz ±5%
Modes: Continuous
Pulsed—20% duty cycle
Pulsed—50% duty cycle
Modulation:
100%
Modulation Waveform:
Rectangular
Pulse Repetition Rate:
Modulation Frequency
100 Hz ±20%
Pulse Duration:
Modulation Period
2 msec ±20%, 20% duty cycle
5 msec ±20%, 50% duty cycle
Temporal Peak/ average
intensity ratio:
5:1 ±20%, 20% duty cycle
2:1 ±20%, 50% duty cycle
Maximum output power:
22 W with a 10 cm² applicator, (ME 7310)
11 W with a 5 cm² applicator, (ME 7513)
2.2 W with a 1 cm² applicator (ME 7331)
Maximum intensity:
2.2 W/cm² with all applicators
Indication accuracy:
±20% (for any level above 10% of maximum)
Output description:
The output waveform is continuous or pulsed as programmed
by the membrane panel control. In the pulse mode the 1, 3.2
or 3.3 MHz square wave pulses are modulated. The power
level is adjusted by varying the pulse amplitude. The pulse
waveforms are shown
below: