Chapter 2: Control Testing
Accu-Chek Customer Care 1-800-858-8072
5.
Check that the Level 1 result is within the
Level 1 range on the test strip vial.
6.
Throw away the used test strip.
If the result is not in the acceptable range,
see Understanding Control Test Results
on page 17.
To keep a control test result from being saved in
memory, call Accu-Chek Customer Care for a
Chek
™
Strip.
16
Chapter 2: Control Testing
Accu-Chek Customer Care 1-800-858-8072
Range (mg/dL)
Level 1 (LO)
42–72
Level 2 (HI)
290–328
Example
Example
Level 2 Testing
Level 1 testing is all you need to do most of the time. If you think your meter or strips may not be working
correctly, you may also want to do a Level 2 test. Simply repeat steps 3 through 6, using Level 2 control
solution.
If you do not have Level 2 control solution, you may purchase control solutions separately.
Understanding Control Test Results
Example
Range (mg/dL)
Level 1 (LO)
42–72
Level 2 (HI)
290–328
The label on your test strip vial shows the acceptable ranges for
Level 1 and Level 2 test solutions. Your control test results should
be within the listed range. Make sure you compare the result to the
range for the level of control solution you used (Level 1 or Level 2).
When a control test result is within the acceptable range,
your test strips and meter are working properly.
“Level 2 Testing...” and step
numbers 5-6 are process of PMS
207 on page 16.
“Understanding Control Test...” is
process of PMS 207 on page 17.
362-17014 Silver Adv OB 3/27/03 5:58 PM Page 16