Teledyne Trace Oxygen Analyzer 3000TA User Manual
Page 18
Quickstart Guide
Trace Oxygen Analyzer
Teledyne Analytical Instruments
18
•
Latching?
Ltch–Y/N
(Yes/No).
•
To change any of the alarm parameters, use the ◄► arrow keys to move the blinking
over to the specific parameter.
•
Use the ▲▼ arrow keys to change the value of the parameter. Holding down the key
speeds up the incrementing or decrementing.
•
Once the parameters for alarm 1 have been set, press Alarms again, and repeat this
procedure for alarm 2 (
AL–2
).
•
To reset a latched alarm, go to
Dft–
and then press either ▲ two times or ▼ two
times. (Toggle it to Y and then back to N.)
—or—
Go to
Ltch–
and then press either ▲ two times or ▼ two times. (Toggle it to N and
back to Y.)
Set The Range Function
To set the 3 programmable ranges:
•
Enter the range function mode by pressing the Range button on the front panel.
L—100 M—1000
H—10000 Mode—AUTO
•
Use the ◄► arrow keys to blink the range to be set: low (L), medium (M), or high
(H).
•
Use the ▲▼ arrow keys to enter the upper value of the range (all ranges begin at 0
ppm). Repeat for each range you want to set.
•
Press Enter to accept the values and return to analyze mode.
Note: The ranges must be increasing from low to high, for example, if range 1 is set as
0–100 ppm and range 2 is set as 0–1,000 ppm, range 3 cannot be set as 0– 500
ppm since it is lower than range 2.
Fixed Range or Autoranging Analysis
To switch from autoranging to fixed range analysis:
•
Enter the range function by pressing the Range button.
•
Use the ◄► arrow keys to move the blinking over
AUTO
.