Teledyne Trace Oxygen Analyzer 3000TA User Manual

Page 18

background image

Quickstart Guide

Trace Oxygen Analyzer


Teledyne Analytical Instruments

18

Latching?

Ltch–Y/N

(Yes/No).

To change any of the alarm parameters, use the ◄► arrow keys to move the blinking
over to the specific parameter.

Use the ▲▼ arrow keys to change the value of the parameter. Holding down the key
speeds up the incrementing or decrementing.

Once the parameters for alarm 1 have been set, press Alarms again, and repeat this
procedure for alarm 2 (

AL–2

).

To reset a latched alarm, go to

Dft–

and then press either ▲ two times or ▼ two

times. (Toggle it to Y and then back to N.)

—or—

Go to

Ltch–

and then press either ▲ two times or ▼ two times. (Toggle it to N and

back to Y.)

Set The Range Function

To set the 3 programmable ranges:

Enter the range function mode by pressing the Range button on the front panel.

L—100 M—1000

H—10000 Mode—AUTO

Use the ◄► arrow keys to blink the range to be set: low (L), medium (M), or high
(H).

Use the ▲▼ arrow keys to enter the upper value of the range (all ranges begin at 0
ppm). Repeat for each range you want to set.

Press Enter to accept the values and return to analyze mode.

Note: The ranges must be increasing from low to high, for example, if range 1 is set as

0–100 ppm and range 2 is set as 0–1,000 ppm, range 3 cannot be set as 0– 500
ppm since it is lower than range 2.

Fixed Range or Autoranging Analysis

To switch from autoranging to fixed range analysis:

Enter the range function by pressing the Range button.

Use the ◄► arrow keys to move the blinking over

AUTO

.