Appendix
A.1 Specifications
Range: 0-100% oxygen
Accuracy: +2% of full scale (at constant temperature
and pressure)
Response
Time: 90% in less than 6 seconds at 25 °C
Battery
Life: Approximately 2000 hours of continuous
use in non-alarm condition
System
Power:
3 AA alkaline batteries.
Sensor
Type:
Class R17MED
Expected Life: 36 months in air at 25°C, Atmospheric
Pressure and 50% Relative Humidity.
(~7.5 months under similar conditions
when continuously exposed to 100% oxy-
gen)
Dimensions: 2.5" W × 1.25" D × 4.5" H (66 mm x 33-
mm × 111.5 mm)
Sensor
Cable: Retracted: 2 ft / Extended: 10 ft.
Storage Temp. 10-30°C (continuous), 5-50°C (Intermit-
tent)
Operating Temp: 0-40°C
Alarm
ranges: HI - 19 to 100%; LO - 18 to 99 %
Alarm accuracy: Alarm thresholds are digitally stored and
alarm affectivity is digitally compared
against the value calculated by the micropro-
cessor. Hence, the accuracy of the alarms is
the same as the accuracy of the unit as speci-
fied above.