1 features – Teledyne MX300-I User Manual
Page 11

Portable Oxygen Monitor
Introduction
Teledyne
Analytical
Instruments
3
NOTE: IF THE LOW ALARM CONDITION IS TRIGGERED WITH
THE OXYGEN LEVEL STAYING ABOVE THE SAFETY LIMIT
OF 18%, THE ALARM BUZZER CAN BE DEACTIVATED BY
PRESSING THE ALARM SILENCE
()
KEY. IF ALARM
CONDITION IS NOT CORRECTED WITHIN ABOUT 115
SECONDS, THE ALARM BUZZER WILL REACTIVATE.
PRESSING THE ALARM SILENCE
()
KEY AGAIN WILL
SILENCE THE ALARM BUZZER ONCE AGAIN FOR ABOUT
115 SECONDS, WITH THE BUZZER REAACTIVATING IF
THE ALARM CONDITIONS PERSIST.
NOTE: IF THE LOW ALARM CONDITION IS TRIGGERED WITH
THE OXYGEN LEVEL FALLING BELOW 18%, THE
√
SENSOR SIGN WILL FLASH ALONG WITH THE GREEN
LIGHT, AND THE BUZZER WILL ACTIVATE. IN THIS CON-
DITION, THE ALARM BUZZER CANNOT BE SILENCED.
WHILE HOLDING DOWN THE ALARM SILENCE
()
KEY
MAY MOMENTARILY SILENCE THE ALARM BUZZER, AUD-
IBLE ALARM WILL RETURN UPON DISENGAGING THE
ALARM SILENCE
()
KEY. HOLDING THE KEY FOR
LONGER DURATIONS WILL INITIATE THE “STUCK KEY”
ALARM, WHICH CAN BE DEACTIVATED ONLY VIA REMOV-
ING THE BATTERIES FROM THE ANALYZER. ANALYZER
SHOULD BE RECALIBRATED UP ON REPOWERING.
1.1 Features
The following features are standard on the MX300-I Monitor:
Large easy to read 3½-digit LCD (see options)
Automatic LCD back lighting upon key press
Microprocessor controlled
Up/Down control keys on the front panel
Sensor fail/disconnect alarm indicator (audible and visual)
Two adjustable concentration alarms with user-defined set
points
Alarm Silence () key
2000 operating hours from 3 AA alkaline batteries
Battery status indicator
Battery test function
Stand for upright tabletop deployment