PRODUCT INSTRUCTION SHEET


Operating temperature range:

0-45 degC

Maximum operating pressure:

1 bar/14.7 psi/1atm

Flow rate minimum:

.20 gpm

Flow rate maximum:

0.6 ppm

pH range:

5.5-9.0

Output signal:

4.0+/- 0.2mA in air (zero)

20mA +/- 0.2mA at high range

(2,

5

or

10ppm)

Power Requirement:

12-24 VDC,

250 mA minimum

Cross-Sensitivity:

ClO2, ozone, bromine

Chemical Compatibility:

up to 50% ethanol/water or

up

to

50%

glycerol/water

OPERATING SPECIFICATIONS

SECTION 12.0

SENSOR SPECIFICATION

12.1 OPERATING SPECIFICATIONS. Follow all operating specifications,
especially for pH and flow rate as noted in the specification tables
below.


SENSOR
Dimensions:

8.2"L x 1" dia

Body Material:

Black PVC

Membrane Material:

polyethersulfone

O-ring material:

Viton

R

Cathode:

Gold

Anode:

Silver chloride (AgCl)

Cable:

2 -conductor shielded, 10ft (3mtr)

tinned

wire

leads

FLOW CELL
Dimensions:

5.58"H x 2.25"DIA

Material:

Acrylic

Connections:

1/4" NPT inlet and outlet

TECHNICAL SPECIFICATIONS

MAINTENANCE/REPLACEMENT PARTS

FC72

Flow Cell, 1/4 inch FNPT inlet and outlet, includes:

2 each 1/4" barbed tube fittings, clamp, threaded

flow cell installation fitting

FCL-ES

Free Chlorine sensor fill solution, 30mL, 1 each

FCL-RM

FCL Replacement premembraned cap, 1 each

FC72 FLOW CELL DIMENSIONS

SENSOR AND FLOW CELL INSTALLATION DIMENSIONS

Parts covered by this product instruction sheet include: FCLTX-100 Series

M4679/0708 page 6 of 6