Product instruction sheet – Omega FCLTX-100 User Manual

Page 4

background image

ZERO

READ

ClO2

C L O 2

to Increase

output

to Decrease
output

SECTION 8.0

CALIBRATION

FIG. 6

SECTION 9.0

SENSOR STORAGE

IMPORTANT NOTE: SENSORS ARE SUPPLIED FACTORY CALIBRATED WITH A
4-20mA SIGNAL OUTPUT CORRESPONDING TO THEIR SPECIFIC RANGE (0-2,
0-5 OR 0-10ppm). SPAN CALIBRATION IS NECESSARY WHEN RECEIVING A
NEW SENSOR SINCE YOUR CONDITIONS MAY VARY FROM THOSE USED AT
THE FACTORY FOR SPAN CALIBRATION. THE ZERO POINT CALIBRATION IS
NOT NECESSARY SINCE THE ZERO SETTING IS VERY STABLE.

PERIODIC CALIBRATION ( ABOUT ONCE PER WEEK) IS RECOMMENDED. THIS
IS USEFUL IN TRACKING SENSOR FAILURES AS WELL.

8.1 Removal of cord grip to access ZERO and SPAN
potentiometers inside the sensor body:

a. Remove top nut of cord grip as shown in FIGURE 4.

b. Remove remaining cord grip as shown in FIGURE 5. A wrench

may be required.

8.2 Slope Calibration:

a. Determine the free chlorine content using a diethyl-p-

phenylenediamine (DPD) colorimeter test kit (see FIGURE 9.),

not included with ClO2 sensor and flow cell.

b. Measure Free Chlorine content with sensor. Make sure that

calibration flow rate and pH matches flow rate when

measuring sample since probe output is flow rate and pH

dependent. Make sure pH is within 5.5-8.0 range.

c. Adjust the span potentiometer to the desired mA (see

FIGURES 9 & 9A)

c. Repeat this slope calibration one day after sensor is

initially installed.

d. Repeat the slope calibration monthly.

9.0 STORAGE. Store sensor at 5

o

C- 50

o

C ONLY and maximum

humidity of 90% (non-condensing).

a. Short Term Storage (1 week or less): Store in Flow cell

with water to prevent the probe from drying out.

b. Intermediate Term (1 week to 1 month): Store in cap,

bottle, or beaker with water to keep membrane wet.

c. Long Term (1 month or longer): Remove Membrane

Cap and store completely immersed in tap water. Turn

sensor upright and shake it to remove fill solution from

inside the sensor.

d. Electrolyte bottle shelf-life = 1 year from date of

manufature(see expiration date on bottle)

PRODUCT INSTRUCTION SHEET

Left
potentiometer
(Zero adjust)

Right
potentiometer
(Span adjust)

Cable

FIG. 7

FIG. 8

FIG. 9

Note: The DPD method
for Free chlorine mea-
surement is not selective
for free chlorine alone.
Other oxidants in the
water can also be mea-
sured as free chlorine.

FIG. 4

FIG. 5

Free Cl

Free Cl

Parts covered by this product instruction sheet include: FCLTX-100 Series