• Before measuring, stretch your back and sit up

straight, breath normally 5-6 times, relax your
shoulders, arms and entire body.

• Slightly bend and support your elbow (e.g. on a

table).

• Do not move, talk or touch the device during

measurement.

• Stress influences your blood pressure, so try to

relax before taking a measurement.

• To compare results, measurements must always

be taken from the same arm.

• Do not place the cuff over thick clothes and do not

roll up your sleeve if it is too tight.

• Align the cuff with the height of your heart (breast

height). If necessary use a pillow or cushion.

• Coldness raises your blood pressure. It is

recommended to measure at room temperature
(+20°C).

• Relax before measurement. If necessary go to the

bathroom before measurement.

• Do not measure directly after bathing.

• If you wish to interrupt measurement prematurely,

press the (O/I) button. Inflation is stopped and the
air is automatically released from the cuff.

• Blood pressure measurements taken in a moving

vehicle may be influenced by vibration.

• If an error has occurred during measurement

is

displayed. Please refer to the next chapter.

• The monitor should be switched off after

measurement to avoid running down the batteries
unnecessarily. If you forget to do so, the monitor
switches itself off automatically after 5 minutes.

E

Hints on taking
blood pressure
readings

Failure, causes and rectification

Failure and their possible causes

Incorrect measurement is indicated by

1. A correct reading could not be obtained

because measurement was disturbed by
movement of the body.

2. The cuff is not fitted correctly.

3. Your clothing has constricted your

blood flow.

4. There is still air in the cuff when the

monitor is switched off.

The (START

) button was pressed

before the (

) symbol was displayed.

The blood pressure values displayed are
extremely low or high, or they are
implausible.

The display does not light up when the
(START

) button is pressed.

1. The batteries are exhausted.

2. The +/- poles of the batteries have been

reversed.

3. The battery contacts are dirty.

The batteries are weak or exhausted.

The cuff pressure does not rise
although the pump motor can be heard.

problem with mem. function

Er

E

Rectification

1. Repeat the measurement keeping

perfectly still. Do not move your arm and
do not speak.

2. Check that the cuff is correctly fitted.

Then repeat the measurement.

3. Remove the item of clothing which

caused the constriction.

4. The unit may be defective. Please have

it examined and, if necessary, repaired
by OMRON.

Wait for the (

) symbol indicating

readiness before you press the (START

)

button.

Refer to the instructions and then repeat
the measurement.

1. Check the batteries and, if necessary, fit

4 new batteries.

2. Reinsert the batteries with the +/- poles

the right way round.

3. Clean the battery contacts with a dry

cloth.

Fit new batteries.

Check that the air tube is properly
connected to the monitor. Push the tube
connector firmly into the socket.

Contact your Omron retailer or distributor.