13. Durability
13.1. Electrical Noise
No malfunction should occur when sinusoidal electrical noise (10Hz -100kHz, < 0.1Vpp)
is added to a power supply line.
Conditions
Barcode Sample: OPTOELECTRONICS Test Sample
PCS: 0.9
Resolution: 0.25
mm
Symbology: 9-digit
Code-39
Quiet Zone:
10 mm
N/W Ratio:
1:2.5
Distance:
100 mm from the edge of scanner
Angle:
α = 0° β = 15° γ = 0°
Curvature: R
=
∞
Power Supply Voltage:
6.0 V (RS-232C) / 5.0 V (USB and Wedge)
13.2. Static Electricity
Air discharge:
8 kV max (No malfunction)
15 kV max (No destruction)
Contact discharge:
6 kV
Contact discharge at the hook and the screw
Measurement environment:
Use electrostatic testing device compliant with IEC 61000-4-2
Discharge resistance:
330 Ω
Capacitor charging:
150 pF