Model:

HEM 712C

Display:

Digital LCD

Measurement Range:

Pressure: 0-280mmHg
Pulse: 40-200/minute

Accuracy:

Pressure: ±3 mmHg
or 2% of the reading
Pulse: ±5% of reading

Inflation:

Automatic by
electric pump

Deflation:

Automatic pressure
release valve

Rapid Pressure Release:

Automatic exhaust valve

Pressure Detection:

Capacitance
pressure sensor

Measurement Method:

Oscillometric method

Pulse wave Detection:

Capacitance
pressuresensor

Power Source:

4 “AA” batteries
(not included)
AC Adapter

Battery Life:

Approx. 300 uses
when used once a
day for 2 minutes
(alkaline battery)

Operating Temperature /
Humidity:

50°F to 104°F
(10°C to 40°C)
30-85% RH maximum

Storage Temperature /
Humidity:

-4°F to140°F
(-20°C to 60°C)
10-95% RH maximum

Console Weight w/
batteries:

Approx. 13 oz (380g)

with batteries

Outer Dimensions:

Approx.
4 1/4" (l) x 5" (w) x 2

1/2"(h)

(107mm x 128mm x 63mm)

Cuff Dimensions:

Approx.

5 1/2"(w) x 18 7/8"(l)

(140mm x 480mm)

Arm Circumference:

9" to 13" (220mm -320mm)

Accessories:

Cuff, Instruction manual

UPC Code

0 73796 07122 6

Except for batteries, NO USER
SERVICEABLE PARTS IN THE
CONSOLE (MAIN UNIT).
NOTE: These specifications,
to improve performance, are
subject to change without notice.

8

ERROR INDICATORS

If you continue to experience difficulty obtaining a reading, call
OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350.
Please tell the representative you have a Omron Model HEM-712C Automatic Inflation
Blood Pressure Monitor.

Error Symbol

Cause

Correction

Cuff under-inflated

.

Carefully read and
repeat in steps listed
under “How To Take
A Reading.”

Batteries have
run down.

Replace all four
“AA”batteries.

Wait 2-3 minutes and
take another measurement.

Repeat steps listed
under “How To
Take A Reading”
but manually inflate
the cuff 40 mmHg
higher than the
previous inflation value.

Movement during

Measurement.

No pulse or blood

pressure reading

due to very weak

artery pulse.

Cuff over-inflated
(more than 300mmHg).

SPECIFICATIONS

9