Troubleshooting guide – Omega 3200ST User Manual
Page 23
M-4271/0707, pg. 23 of 26
H.
Troubleshooting Guide
Symptom
Possible Cause
Method of Correction
Clicking noise from
controller
Unit in error mode
Check there is sufficient
pressure and that the flow
path is not restricted or
blocked
Unit wired incorrectly
Check wiring is according to
Section B5
Loose connection
Check all connectors and
wiring
Damaged connector pins
Contact Omega Engineering
Blocked flow path
Check flow path for
obstructions.
Piping leak before sensor
Check all piping and
connections.
Insufficient power
Check the power supply
output and increase if
necessary
Output load resistance too low
Ensure the voltmeter or
data acquisition system or
display has an impedance of
greater than of 2.5kohm
Flow too low for the unit
Ensure that the flow being
measured is within the
capabilities of the unit
No response
Unit damaged or faulty
Contact Omega Engineering
Particles in flow path
Add filtration before the
sensor.
Flow path obscured
Remove any debris or
blockage in the flow path
eg. PTFE tape.
Unit calibrated for a different
gas
Check calibration certificate
and apply a “K” Factor to
readings if necessary.
Gas composition is variable
Contact Omega Engineering
Fittings have been changed
Replace the factory installed
fittings
Moisture in gas
Ensure gas is clean and dry
Insufficient warm-up period
Allow the unit to warm-up
for at least 5 minutes.
Zero drift
Verify the zero and adjust
as necessary as explained in
Section C
The gain potentiometer has
been adjusted
Contact Omega Engineering
Unit needs recalibration
Contact Omega Engineering
Inaccurate control
Flow too high for the unit
Ensure that the flow being
measured is within the
capabilities of the unit