Model:

OMRON M5-I

Display:

LCD digital display

Blood Pressure

Pressure: 40 mm Hg to 280 mm Hg

Measurement Range:

Pulse: 40 – 180/min.

Accuracy:

Pressure: ±3 mm Hg
Pulse: ±5% of reading

Inflation:

Fuzzy-Logic controlled by electric pump

Deflation:

Automatic pressure release valve

Pressure Detection:

Capacitive pressure sensor

Measurement Method:

Oscillometric method

Power Source:

4 Alkaline batteries 1.5V (Type LR6)

or AC/DC adapter (optional, 6V = 4W)

Battery Life:

Capacity of new batteries is approx.
300 measurements

Operating Temperature:

10°C to 40°C

Storage Temperature:

-20°C to 60°C

Humidity:

30 to 85% RH

Storage humidity:

10 to 95% RH

Console Weight:

Approximately 530g with batteries

Outer Dimensions:

Approximately 115 mm x 177 mm x 72 mm

Cuff Dimensions:

Approximately 140 mm x 480 mm
( arm circumference 22 - 32 cm )

Accessories:

Cuff, Type M (CR17); instruction manual,
storage case, battery set

Manufacturer:

OMRON Matsusaka Co. Ltd., Japan

NOTE: Subject to technical modification without prior notice

English

10

11

M5-I

Technical data

This device fulfils the provisions of the EC directive
93/42/EEC (Medical Device Directive). This blood
pressure monitor is designed according to the
European standard EN1060, Non-invasive sphyg-
momanometers Part 1: General requirements and
Part 3: Supplementary requirements for electrome-
chanical blood pressure measuring systems.

• For cleaning the monitor only use a soft, lightly moi-

stened cloth.
Do not use petroleum spirits, thinners or similar
solvents!

• Stains on the cuff can be carefully removed using a

moist cloth and soapy water.
Do not wash the cuff or make it wet!

• Keep your blood pressure monitor in the storage

case to protect it from dust and moisture.

• Protect your monitor against vibrations and do not

leave it in a place where temperatures are extreme-
ly low (below -20 °C) or extremely high (above 60
°C) or with high humidity(+ 85%).

• Do not use your monitor at very low temperatures

(below 10 °C) or very high temperatures (above 40
°C).

• Do not fold the cuff or air tube together too tightly

and do not crease them.

• Do not carry out repairs of any kind yourself.

If a defect occurs, consult the OMRON Customer
Service or seek the advice of your surgical applian-
ce stockist or pharmacist.

• The well confirmed accuracy of this blood

pressure monitor is designed to last a long life.

• In order to increase the lifetime of the batteries

take them out when you store the device for a
longer period (2 months).

• Electro-magnetic fields can influence the measu-

rement. Do not use a potable phone near this unit

Maintenance and
storage

• Disposal of this product and used batteries should be carried out in accordance

with the national regulations for the disposal of electronic products.

= Type B