M5-i, Failure, causes and rectification, Hints – Omron Healthcare M5-I User Manual

Page 5: English

background image

English

8

9

M5-I

Failure and their possible causes

Incorrect measurement is indicated by E

1. A correct reading could not be obtained

because measurement was disturbed by
movement of the body.

2. The cuff is not fitted correctly.

3. Your clothing has constricted your blood flow.

4. There is still air in the cuff when the monitor

is switched on.

The start button was pressed before the
(

) symbol was displayed

The blood pressure values displayed are ex-
tremely low or high, or they are implausible.

The display does not light up when the (

)

button is pressed.

1. The batteries are exhausted.

2. The +/- poles of the batteries have been

reversed.

3. The battery contacts are dirty.

The batteries are weak or exhausted.

The cuff pressure does not rise although
the pump motor can be heard.

Err

problem with mem. function

Failure, causes and rectification

Rectification

1. Repeat the measurement keeping

perfectly still. Do not move your arm and
do not speak.

2. Check that the cuff is correctly fitted. Then

repeat the measurement.

3. Remove the item of clothing which cau-

sed the constriction.

4. The unit may be defective. Please have it

examined and, if necessary, repaired by
OMRON Customer Services.

Wait for the (

) symbol indicating readi-

ness before you press the start button.

Refer to the instructions and then repeat the
measurement.

1. Check the batteries and, if necessary, fit 4

new batteries.

2. Reinsert the batteries with the +/- poles

the right way round.

3. Clean the battery contacts with a dry cloth.

Fit new batteries.

Check that the air tube is properly connec-
ted to the monitor. Push the tube connector
firmly into the socket.

Contact Omron Customer Service.

• Before measuring, stretch your back and sit up straight,

breathe normally 5-6 times, relax your shoulders and
arms and entire body.

• Lightly bend and find support for your elbow ( e.g. on the

table ).

• Do not move, talk or touch the device during measure-

ment.

• Stress influences your blood pressure. Relax before

measurement.

• To compare results measurements must always be

taken from the same arm.

• Do not place the cuff over thick clothes and do not roll up

your sleeve if it is too tight.

• Align the cuff with the height of your heart (breast

height). If necessary use a pillow or cushion.

• Coldness raises your blood pressure. Recommended is

measuring at room temperature ( ±20°C).

• Relax before measurement. If necessary go to the bath-

room before measurement.

• Do not measure directly after bathing.
• When measuring in a moving vehicle the measurement

can be influenced.

• If you wish to interrupt measurement prematurely, press

the (

) button. Inflation is stopped and the air is auto-

matically released from the cuff.

• If you wish to repeat the measurement, follow the same

procedure as before starting at step 4 (Inflating the cuff).

• If an error has occurred during measurement E is

displayed. Please refer to the next chapter.

• The monitor should be switched off after measurement

to avoid running down the batteries unnecessarily. If you
forget to do so, the monitor switches itself off automati-
cally after 5 minutes.

Hints