8. Technical

Specifications

The conditions for technical specifications are as follows, unless otherwise specified in each
section.

Conditions

Ambient temperature and humidity

Room temperature (5–35° C) and room humidity
(45% RH–85% RH)

Ambient light

500 to 900 lx

Background

Barcode = black
Space = white
Margin = white
Background of label = black

8.1. Print Contrast Signal (PCS)

0.45 or higher (over 70% of reflectivity of space and quiet zone).

PCS is determined by: Reflectance of white bar – Reflectance of black bar / Reflectance
of white bar

8.2. Scan Area and Resolution

The depth of field is measured from the edge of the data collector. The scanning range is
within the circular arc centered on the scan origin.