Technical Specifications

Pressure–Flow Curve

Note: As a result of manufacturing variations, the vent flow may vary from the
values listed in the table.

Dead Space Information

Dead space is the empty volume of the pillow sleeve and short tube assembly

up to the swivel.

The dead space of the Mirage Swift II for all pillow sleeve sizes is 91 mL.

Standards

CE designation in accordance with EC directive 93/42/EEC, class IIa.

EN ISO 14971, ISO 5356-1, ISO 10993-1 and EN ISO 17510-2.

Therapy Pressure

4 to 20 cm H

2

O

Resistance

Drop in pressure measured (average for three pillow sleeve sizes)

at 50 L/min: 0.4 cm H

2

O

at 100 L/min: 1.6 cm H

2

O