beautypg.com

National Instruments 1192 User Manual

Page 20

background image

Chapter 1

Installing and Configuring the SCXI-119X

SCXI-1190/1191/1192 User Manual

1-12

ni.com

e.

Change the Connected to selection from None to Connected to:
Device x
.

If you get the warning message

Failed to find

followed by the

module codes and the message

Unable to communicate with

chassis

, take the following troubleshooting actions:

a.

Make sure that the SCXI chassis is powered on.

b.

Make sure the cable between the SCXI chassis and DAQ device or
computer-based instrument is properly connected.

c.

Inspect the cable and the connectors for any bent pins.

d.

Make sure you are using the correct National Instruments cable
assembly.

e.

Test your DAQ device or computer-based instrument to verify it is
working properly. Refer to your DAQ device or computer-based
instrument user manual for more information.

If you get the warning message

Failed to find

, followed by

module codes and then the message

Instead found: module with

ID 0Xxx

:

a.

Return to the

Connecting the SCXI-119X to a DAQ Device or

Computer-Based Instrument

section and make sure the correct

module is in the specified slot.

b.

Delete the incorrect module as described in the

Removing the

SCXI-119X from an SCXI Chassis

section.

c.

Add the correct module as described in the

Manually Adding

Modules

section.

If you get the warning message

Failed to find

, followed by a

module code and then the message

Slot x is empty

:

a.

Check to see if the configured module is installed in the specified
slot. If not, install the module by referring to the

Installing the

Hardware

section.

b.

If the module is installed in the correct slot, turn off the chassis,
and remove the module as specified in the

Removing the

SCXI-119X from an SCXI Chassis

section. Verify that no

connector pins are bent on the rear signal connector. Reinstall the
module, ensuring the module is fully inserted and properly aligned
in the slot.

After checking the preceding items, return to the

Self-Test Verification

section and retest your SCXI chassis.