National Instruments NI 5412 User Manual
Page 3
© National Instruments Corporation
3
NI 5412 Specifications
Maximum Output Voltage
Maximum
Output
Voltage
Path
Load
Maximum Output Voltage (V
pk-pk
)
The Maximum
Output Voltage
of the NI 5412 is
determined by
the Amplitude
Range and the
Offset Range.
Low-
Gain
Amplifier
50
Ω
±1.000
1 k
Ω
±1.905
Open
±2.000
High-
Gain
Amplifier
50
Ω
±6.000
1 k
Ω
±11.43
Open
±12.00
Accuracy
DC Accuracy
±0.2% of Amplitude ± 0.05% of Offset ± 500 µV
(within ±10 °C of self-calibration temperature)
±0.4% of Amplitude ± 0.05% of Offset ± 1 mV
(0 °C to 55 °C)
Calibrated for
high impedance
load.
AC Amplitude
Accuracy
±1.0% of desired Amplitude ± 1 mV
50 kHz sine
wave.
Output Characteristics
Output
Impedance
50
Ω or 75 Ω nominal, software-selectable.
—
Output
Coupling
DC
—
Output Enable
Software-selectable. When the Output Path is disabled, the
CH 0 Output is terminated to ground with a 1 W resistor
equal to the selected output impedance.
—
Maximum
Output
Overload
The CH 0 output can be connected to a 50
Ω, ±12 V source
without sustaining any damage. No damage occurs if the
CH 0 output is shorted to ground indefinitely.
—
Waveform
Summing
The CH 0 output supports waveform summing among
similar paths—specifically, the outputs of multiple NI 5412
signal generators can be connected directly together.
—
Table 1. (Continued)
Specification
Value
Comments