ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):
1. Indoor use;
2. Altitude up to 2000 m;
3. Ambient temperature 5
o
C to 40
o
C
4. Maximum relative humidity 80% for temperature up to 31
o
C decreasing linearly to 50% relative
humidity at 40
o
C;
5. Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;
6. Other supply voltage fluctuations as stated by the manufacture;
7. Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains
supply the minimum and normal category is II;
8. Pollution degree 2 in accordance with IEC 60664.