Auto cal menu – Sony STR-DH810 User Manual
Page 40

40
GB
*
xxx
represent a speaker channel
FL
Front left
FR
Front right
CNT
Center
SL
Surround left
SR
Surround right
SBL
Surround back left
SBR
Surround back right
LH
Front left high
RH
Front right high
SW
Subwoofer
Depending on the measurement result, the speaker
channel may not appears.
To return to step 2 of
“3: Confirming/saving the
measurement results”
Press .
Tip
Depending on the position of the subwoofer, the
measurement results for polarity may vary.
However, there will be no problems even if you
continue to use the receiver with that value.
You can use the AUTO CAL menu to make
various adjustments for Auto Calibration
settings and to name the position.
For details on adjusting the parameters, see
“Using the setting menu” (page 72).
x A.CAL START
x CAL TYPE*
Lets you select the Auto Calibration type for
each seating position. For details, see step 4 of
“3: Confirming/saving the measurement
results” (page 38).
* You can select this parameter only when you have
performed the Auto Calibration and saved the
settings.
x POSITION
You can register three patterns as position 1, 2
and 3, depending on the seating positions,
listening environment, and measurement
conditions.
If you do not select the position number, the
Auto Calibration result will automatically save
as POS 1 (initial setting).
To register settings for the
listening environment
You can select the seating position you want
and register the measurement results of the
Auto Calibration for that seating position.
1
Press AMP MENU.
2
Press
V/v repeatedly to select “ CAL>”, then press or b. 3 Press V/v repeatedly to select “POSITION”, then press or b. 4 Press V/v repeatedly to select the position (POS 1, 2, 3) for which you . W - xxx * : 42 The volume of the receiver is out of the acceptable W - xxx * : 43 The distance and position of a subwoofer cannot be NO WARN Display and explanation AUTO CAL menu
want to register the measurement
results, then press
range. Try to perform the measurement when the
environment is quiet enough to allow proper
measurement.
detected. This may be caused by noise. Try to
perform the measurement in a quiet environment.
There is no warning information.