Page 23

background image

22

ALARMS & SAFETY FUNCTIONS

This unit has the alarms and safety functions shown in Table 5, and also self diagnostic functions.

Table 5 Alarms and safety functions

Alarm & Safety

Situation

Indication

Buzzer

Safety operation

High temperature
alarm

If the chamber temperature is higher
than the temperature at which the
high temperature alarm is activated.

ALARM lamp is flashed.
Temperature indicator is
flashed.

Intermittent tone with
15 minutes delay.

Remote alarm with 15
minutes delay.

Low temperature
alarm

If the chamber temperature is lower
than the temperature at which the
low temperature alarm is activated.

ALARM lamp is flashed.
Temperature indicator is
flashed.

Intermittent tone with
15 minutes delay.

Remote alarm with 15
minutes delay.

Power failure alarm

In the case of power failure.
When power switch is turned OFF.
When the power to the unit is
disconnected.

ALARM lamp is flashed.

Intermittent tone

Remote alarm.

Auto-return

When there is no key pressing in
each setting mode for 90 seconds.

Chamber temperature is
displayed.

-----

Finishing of each
setting mode.

Key lock

When the key lock is “ON”.

-----

-----

Change of setting is
disable.

If the thermal sensor is
disconnected.

ALARM lamp is flashed.
E01 and chamber temp. are
displayed alternately.

Intermittent tone

Remote alarm.
Continuous running.

Thermal sensor
Abnormality
(for MDF-U537D,
upper chamber
only)

If the thermal sensor is
short-circuited.

ALARM lamp is flashed.
E02 and chamber temp. are
displayed alternately.

Intermittent tone

Remote alarm.
Continuous running.

If the thermal sensor is
disconnected.

ALARM lamp is flashed.
E03 and chamber temp. are
displayed alternately.

Intermittent tone

Remote alarm.
Continuous running.

Thermal sensor
abnormality
(MDF-U537D,
lower chamber
only)

If the thermal sensor is
short-circuited.

ALARM lamp is flashed.
E04 and chamber temp. are
displayed alternately.

Intermittent tone

Remote alarm.
Continuous running.

If the protective sensor for
compressor is disconnected.

ALARM lamp is flashed.
E05 and chamber temp. are
displayed alternately.

Thermal sensor
abnormality
(MDF-U537, U537D
only)

If the protective sensor for
compressor is short-circuited.

ALARM lamp is flashed.
E06 and chamber temp. are
displayed alternately.

Intermittent tone

Remote alarm.
Normal operation.

If the thermal sensor for lower
chamber is disconnected.

ALARM lamp is flashed.
E07 and chamber temp. are
displayed alternately.

Thermal sensor
abnormality
(MDF-U537 only)

If the thermal sensor for lower
chamber is short-circuited.

ALARM lamp is flashed.
E08 and chamber temp. are
displayed alternately.

Intermittent tone

Remote alarm.
Normal operation.

Battery switch

check

When battery switch is OFF at the
time of alarm test.

ALARM lamp is flashed.
E09 is flashed.

Intermittent tone

Remote alarm.

Compressor temp.
Abnormality (MDF-
U537, U537D only)

In the case of failure of compressor
cooling fan motor.
In the case of abnormal high
temperature due to the dust on the
condenser.
In the case of abnormal high
ambient temperature.

E10 and chamber temp. are
displayed alternately.

Intermittent tone

Remote alarm.
Compressor stops
running when the temp
reaches about 100

o

C.

Electromagnetic valve
1, 2 and compressor
stop.
(MDF-U537D only)

Note:

1. The above power failure alarm is available when the battery switch is ON and the battery is charged.

If the battery switch is OFF or the battery is discharged, only the remote alarm is activated.

2. The power failure alarm can be kept about 12 hours with the battery charged completely. 2-day

operation of the freezer is needed to charge the battery full.

3. The chamber temperature is displayed for 5 seconds if the BUZZER key is depressed during the power

failure alarm. At the same time, the alarm stops.

4. The remote alarm is silenced by pressing BUZZER key as the remote alarm is operated in conjunction

with the buzzer, except for the power failure alarm.

5. After power failure, the operation is resumed with the condition before power failure since the

temperature setting and alarm temperature setting are memorized in a nonvolatile memory.