JAI IIT Camera Systems VIS-CAM 400 User Manual
Page 40

IIT Camera System Manual
40
Appendix F
Besides these constants there are some commands for test purposes. These are:
Name Cmd
Syntax
Description
Disable checksum
U
$u99
Disables checksum commands until next
power down
Firmware version
V
$V
Shows the installed firmware version for U1
Status request
S
$S
The format of the answer is
$S
stauts>
The temperature is shown in two’s
complement.
Example of an answer:
$S1A800
“1A” is 26 deg
“8” is power ok
“0” fan off, no current, no overload
“0” heater off, no current, no overload
Heater and fan status:
0: off, no current, no overload (ok)
1: on, no current, no overload (fault)
2: off, no current, no overload (fault)
3: on, no current, no overload (ok)
4: off, no current, no overload (fault)
5: on, no current, no overload (fault)
6: off, current, overload (fault)
7: on, current, overload (fault)