JAI IIT Camera Systems VIS-CAM 400 User Manual
Page 31

 
 
System Set-Up
31
IIT Camera System Manual
For more details on FTP image transfer, see section 3.2 of User’s Guide EN Camera Series
Caution: ENSetup Capture Mode and FTP Transfer Mode are exclusive to each other. Both modes
should not be enabled simultaneously in the camera.
4.4 Setting the IIT I/O Board internal trigger parameters
A number of variables can be setup in the IIT I/O Board for optimizing performance. These are 
illustrated in the block diagram in figure 19: IIT Camera Trigger Generation Principle. 
Commands can be launched to the board from the EN Setup application Video window by right 
clicking the mouse in the left side a selecting Serial Terminal and then Laser Trigger (IIT I/O Board). 
Figure 29. Serial Communication Terminal
All variables are set as hexadecimal numbers. All commands are terminated by a 
return). Parameters are always two digits (01 to FF). 
The commands are normally protected by a checksum that can be disabled by launching the 
command #u98. The answer will be #u. 
All commands are hereafter the pound key (#) followed by a capital command letter. The current 
value of the variable is returned if the command is without parameter. 
Example: Command #C