INVACARE 8000 PRESSURE MATTRESS

Part No. 1078996 Rev. B

FIGURE 4 - SECURING MATTRESS BASE

TO BED FRAME

1. Wrap one (1) of the ten (10) straps on

the mattress around the bed frame.

2. Insert strap into the two (2) double D-

rings.

3. Feed strap back through only one (1) of

the double D-rings.

4. Pull the strap tight to secure the mattress

base to the bed frame.

5. Repeat STEPS 1 - 4 for the remaining

nine (9) double D-rings.

NOTE: Each SIDE contains three (3) double D-rings

and each END contains two (2) double D-rings for

a total of ten (10) double D-rings.
6. Adjust the mattress to the proper

firmness. Refer to PLACING

PATIENT ON MATTRESS AND

ADJUSTING FIRMNESS in this

instruction sheet.

NOTE: Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet.
1. Place patient on the mattress.
2. Check the level that the patient sinks into mattress.
NOTE: The pressure of the power unit can be adjusted to provide maximum effectiveness and

pressure relief for the patient.
NOTE: The pressure can be adjusted with a range from -MIN to +MAX.
NOTE: Every other cushion inside the mattress is connected together known as a chamber. The

mattress consists of two (2) chambers.
NOTE: The power unit will cycle every eight (8) minutes. It will alternate inflating and deflating the

two (2) air chambers.
NOTE: The patients body should sink approximately 1/3 into the mattress.
3. If the level is unacceptable, do one (1) of the following:

A. To increase firmness - turn knob towards +MAX setting.
B. To decrease firmness - turn knob towards -MIN setting.

NOTE: After adjusting the knob, wait ten (10) minutes before rechecking the level the patient sinks

into mattress.
4. Repeat STEP 3 until the patients body sinks 1/3 into the mattress.

SECURING MATTRESS BASE TO BED FRAME (FIGURE 4)

Double D-rings

Strap

PLACING PATIENT ON MATTRESS AND ADJUSTING

FIRMNESS (FIGURE 5)