Troubleshooting display codes – Invacare RLS6 User Manual
Page 7

Part No 1114832
7
Reliant Scale RLS6
7. Press the LOCK/UNLOCK key. The display reads UNLD.
8. Press the ZERO key. The display reads LOAD.
9. Hang 200 pounds of calibrated weight from the scale and press the ZERO key.
10. Press the ZERO key. The display reads DONE.
11. Remove the weight. Press the LOCK/UNLOCK key.
NOTE: The scale is now in test mode.
12. Press the ZERO key. The display reads 0.00 lb.
NOTE: The scale will now be in normal operation.
FIGURE 5 Calibrate the Reliant Scale
Troubleshooting
Display Codes
Calibration Required ‐ Indicates improper stored calibration data, calibration is necessary.
Over Capacity ‐ Indicates a weight exceeding the capacity has been loaded on the scale.
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
Unit does NOT work properly.
Battery failure.
Check battery. Replace if necessary.
Battery has been replaced and unit
still does NOT work properly.
Contact Invacare for Service at
1-800-333-6900.
CAL Button
BACK VIEW OF THE SCALE DISPLAY
FRONT VIEW OF SCALE DISPLAY
Display Window