SECTION 5—MAINTENANCE
27
Heavy Duty Cleaning Procedure
ƽ WARNING
DO NOT use this product or any available optional equipment without first completely reading
and understanding these instructions and any additional instructional material such as owner’s
manuals, service manuals or instruction sheets supplied with this product or optional
equipment. If you are unable to understand the warnings, cautions or instructions, contact a
healthcare professional, dealer or technical personnel before attempting to use this equipment
- otherwise, injury or damage may occur.
The disinfectant dispensing system has been factory tested and calibrated for use with Invacare
disinfectant. The use of any disinfectant with a dilution ratio (½ oz per gallon) or viscosity level
different than that of Invacare disinfectant will result in an improper mixture of disinfectant and
water. This may cause your system to be improperly disinfected - or in some instances, not
disinfected at all. If using any disinfectant with a dilution ratio or viscosity level different than
that of Invacare disinfectant or if switching from one type of disinfectant to another type, the
disinfectant flowmeter must be calibrated before performing this procedure. Refer to the tub
owner’s manual or to the calibration instruction sheet, part number 1154272, for the calibration
procedure.
ALWAYS wear rubber gloves, an apron and face shield protection during disinfecting,
cleaning and rinsing procedures. In case of eye or skin contact, follow procedures on
disinfectant container.
CAUTION
NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice
stone. The use of these types of cleaners will make the gel-coat finish of your tub scratched
and dull. NEVER use cleaners containing iodine, bromine, Betadine or methylene blue. The
use of these types of cleaners will cause the tub to stain. Never use cleaners containing
bleach. Use of these types of cleaners will dry out the rubber seals and gaskets and the tub
will not function properly.
Before starting the whirlpool pump, the water level must be completely above the whirlpool
jets. Otherwise damage to the motor/pump will result.
In addition to disinfecting your tub after every use, you will need to perform this procedure
approximately once every 80‐150 baths.
1. Clean the entire surface of the tub to remove visible tissue residue and fluid.
2. Rinse the inside surface of the tub with the shower/massage wand on the spray setting.
3. Rotate the water temperature valve to adjust the water temperature to 105
o
F.
NOTE: This will maximize the cleaning solutionʹs ability to clean and disinfect the system.
4. Close the drain.
5. Make sure the whirlpool jets are open.
NOTE: The required amount of water to cover the whirlpool jets is 50 gallons.
6. Fill the tub above the whirlpool jets.
NOTE: The required amount of Invacare Heavy Duty Cleaner is 200 ounces.
7. Add the Invacare Heavy Duty Cleaner to the water in the tub.
8. Spray pre‐mixed Invacare Heavy Duty Cleaner on the remaining surfaces.
9. Press the whirlpool on/off button to turn on the whirlpool pump and let the pump run for 60
seconds.
10. Turn off the pump and allow the heavy duty cleaner/water solution to remain in the tub for 10
minutes.
11. During this ten‐minute waiting time, take a long handled brush and thoroughly scrub all
interior surfaces of the tub.
12. Open the drain to allow the heavy duty/water solution to drain from the tub.
13. Use the shower/massage wand on the pulse setting and direct the spray into both whirlpool
jets to back flush until the water discharging from the whirlpool inlet is clear.
14. Rinse the tub thoroughly with the shower wand on the spray setting.
15. Check the tub for damage and to ensure all surfaces of the tub are clean and that no visible
residuals remain. If residuals are found, repeat STEPS 13 and 14.