For future servicing, please
document any alterations to the
standard factory settings.

General note

Program selector positions that have
not been allocated to a function are
indicated in the middle part of the
display by a bar (

-).

The programming level is called up via
the "Stop"

f position.

Accessing the programming
level

Before performing any of the special
functions that follow, the programming
mode of the labwasher controller needs
to be accessed.

^ Turn the program selector to "Stop"

f.

^ Turn the machine off.
^ Press and hold the "Display" ß and

"Start"

6 buttons. At the same time,

turn the machine on with the "On/Off"
button.

^ The display will show the current

program status "P...".
The "Fill/Drain" indicator will
illuminate.

^ Press 0 twice. "E02" (programming

level 2) appears in the display.

Programming special functions