Cloning a transmitter – Fluke HART 744 User Manual
Page 27

Documenting Process Calibrator HART Mode
Cloning a Transmitter
19
Cloning a Transmitter
Cloning copies the Basic Setup information from one
transmitter to another. You can clone generic and
supported transmitters.
To use the calibrator to clone a transmitter:
1. Connect the calibrator to the transmitter you want to
clone. You only need a communication connection.
2. From the Active Device screen, press the
Setup
softkey.
3. Use
the
u
or
d
keys to select
Basic
from the list on
the screen and press
e
.
pe09s.bmp
4. Press
the
Store Page
softkey to load the settings into
the calibrator’s memory (not the transmitter).
5. Disconnect the calibrator from the transmitter and
connect it to the transmitter you want to configure.
Press the
Abort
softkey to establish communication.
6. Press the
Setup
softkey, then select
Basic
to return to
the Basic Setup screen.
7. Press the
Recall Page
softkey to refresh the calibrator
display with the cloned parameter settings. The
settings are not transmitted to the transmitter at this
time.
8. Use the
u
or
d
and
e
keys to select
parameters one at a time, change them, or leave them
as cloned, and press the
Send
softkey to transmit
them to the transmitter.