Sensor, Device identification – Fluke HART 744 User Manual
Page 18

744
HART Mode Users Guide
10
Sensor
From the Active Device screen press the
Setup
and
Sensor
softkeys to access the read-only Sensor Setup
screen. This is where you can view information about the
sensor in the transmitter, including serial number, limits,
and span. The limits shown are the absolute limits for the
sensor. (The Upper Range Value (URV) and Lower
Range Value (LRV) are different, and are viewable and
programmable through the Basic Setup screen.)
pe10s.bmp
Device Identification
From the Active Device screen press the
Setup
and
Device Identification
softkeys to view information about the
transmitter. You can program the Tag, Message, Date,
and Descriptor registers in the transmitter using this
screen.
pe11s.bmp