Accuracy specifications – Fluke 115 User Manual
Page 4

114, 115, 116, and 117
Calibration Information
4
Accuracy Specifications
Function Range
Resolution
Accuracy
±
([% of Reading] + [Counts])
Model
[1]
DC millivolts
600.0 mV
0.1 mV
0.5 % + 2
114, 115, 116, 117
DC Volts
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 2
114, 115, 116, 117
DC, 45 to 500 Hz
500 Hz to 1 kHz
Auto-V LoZ
[2]
True-rms
600.0 V
0.1 V
2.0 % + 3
4.0 % + 3
114, 116, 117
45 to 500 Hz
500 Hz to 1 kHz
AC millivolts
[2]
True-rms
600.0 mV
0.1 mV
1.0 % + 3
2.0 % + 3
114, 115, 116, 117
AC Volts
[2]
True-rms
6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3
2.0 % + 3
114, 115, 116, 117
Continuity 600
Ω
1
Ω
Beeper on <20
Ω
, off >250
Ω
; detects opens
or shorts of 500
μ
s or longer.
114, 115, 116, 117
Ohms
600.0
Ω
6.000 k
Ω
60.00 k
Ω
600.0 k
Ω
6.000 M
Ω
40.00 M
Ω
0.1
Ω
0.001 k
Ω
0.01 k
Ω
0.1 k
Ω
0.001 M
Ω
0.01 M
Ω
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 2
114, 115, 116, 117
Diode test
2.000 V
0.001 V
0.9 % + 2
115, 116, 117
Capacitance 1000
nF
10.00
μ
F
100.0
μ
F
9999
μ
F
1 nF
0.01
μ
F
0.1
μ
F
1
μ
F
1.9 % + 2
1.9 % + 2
1.9 % + 2
100
μ
F - 1000
μ
F: 1.9 % + 2
>1000
μ
F: 5 % + 20
115, 116, 117
Lo-Z Capacitance (Power-
up option)
1 nF to 500
μ
F
20% + 2
(10% +2 typical)
115, 116, 117
AC Amps True-rms
[2]
(45 Hz to 500 Hz)
6.000 A
10.00 A
[4]
20 A overload for 30
seconds maximum, 10
minutes rest minimum.
0.001 A
0.01 A
1.5 % + 3
115, 117
DC Amps
6.000 A
10.00 A
[4]
20 A overload for 30
seconds maximum, 10
minutes rest minimum.
0.001 A
0.01 A
1.0 % + 3
115, 117
Temperature (Type K
thermocouple)
-40
°
C to 400
°
C
-40
°
F to 752
°
F
0.1
°
C
0.2
°
F
1.0 % + 10
[5]
1.0 % + 18
[5]
116
AC
μ
Amps True-rms
[2]
(45 Hz to 1 kHz)
600.0
μ
A 0.1
μ
A
1.5 % + 3 (2.5 % + 3 > 500 Hz)
116
DC
μ
Amps 600.0
μ
A 0.1
μ
A
1.0 % + 2
116