Fluke 115 User Manual
Page 15
Digital Multimeters
Calibration Adjustment
15
Notes
Set the calibrator to Standby prior to changing the function switch position and after
completing adjustment of each function.
If the calibration adjustment procedure is not properly completed, the Meter will not operate
correctly.
Table 4. Calibration Adjustment Steps
Calibration Steps
Rotary
Switch
Position
114 115
[1]
116
[1]
117
[1,2]
Input
Terminals
Calibrator
Source Value
e
Ohms
N/A
C/01
C/01
C/01
[2]
No leads
No leads
C/01
C/02
C/02
C/02
V
Ω
/+ and COM
0 V, 0 Hz
C/02
C/03
C/03
C/03
V
Ω
/+ and COM
300 mV, 0 Hz
C/03
C/04
C/04
C/04
V
Ω
/+ and COM
100 mV, 0 Hz
C/04
C/05
C/05
C/05
V
Ω
/+ and COM
-300 mV, 0 Hz
C/05
C/06
C/06
C/06
V
Ω
/+ and COM
60 mV, 0 Hz
C/06
C/07
C/07
C/07
V
Ω
/+ and COM
600 mV, 0 Hz
m
C/07
C/08
C/08
C/08
V
Ω
/+ and COM
600 mV, 60 Hz
C/08
C/09
C/09
C/09
V
Ω
/+ and COM
600
e
, 2-wire comp
C/09
C/10
C/10
C/10
V
Ω
/+ and COM
6 k
e
C/10
C/11
C/11
C/11
V
Ω
/+ and COM
60 k
e
C/11
C/12
C/12
C/12
V
Ω
/+ and COM
600 k
e
C/12
C/13
C/13
C/13
V
Ω
/+ and COM
6 M
e
[3]
C/13
C/14
C/14
C/14
V
Ω
/+ and COM
Short
[3]
e
Ohms
C/14
C/15
C/15
C/15
V
Ω
/+ and COM
40 M
e
[3]
C/15
C/16
C/16
C/16
V
Ω
/+ and COM
6 V, 60 Hz
C/16
C/17
C/17
C/17
V
Ω
/+ and COM
60 V, 60 Hz
C/17
C/18
C/18
C/18
V
Ω
/+ and COM
600 V, 60 Hz
C/18
C/19
C/19
C/19
V
Ω
/+ and COM
6 V, 0 Hz
C/19
C/20
C/20
C/20
V
Ω
/+ and COM
60 V, 0 Hz
K
C/20
C/21
C/21
C/21
V
Ω
/+ and COM
600 V, 0 Hz
Set calibrator to standby, reconfigure leads, and program for amps output.
?
N/A
C/22
N/A
C/22
A and COM
6 A, 60 Hz
[3]
A
N/A
C/23
N/A
C/23
A and COM
6 A, 0 Hz
$
AC
μ
amps
N/A N/A
C/22
N/A
+ and COM
600
μ
A, 60 Hz