Fluke 115 User Manual
Page 10

114, 115, 116, and 117
Calibration Information
10
Table 3. DMM Performance Tests (cont)
Display Reading
Step Function Range Applied
114
[1]
115
[1]
116
[1]
117
[1]
18
6.000V
0V
-0.002 to 0.002
19
6.000V
5V
4.973 to 5.027
20
60.00V
50V
49.73 to 50.27
21
600.0V
600V
596.8 to 603.2
22
L
DC Volts
600.0V
-600V
-596.8 to -603.2
23
600.0 mV
6 mV, 45 Hz
5.6 to 6.4
24
m
AC Millivolts
600.0 mV
600mV, 1 kHz
587.7 to 612.3
25
600.0 mV
10 mV
9.7 to 10.3
26
•
DC Millivolts
600.0 mV
600mV
596.8 to 603.2
27
G
Diode
2.000 V
1.9V
N/A
1.881 to 1.919
28 1000
nF
Open
N/A
0 to 2
0 to 2
0 to 2
29
E
Capacitance
9999
μ
F 900
μ
F
N/A
881 to 919
881 to 919
881 to 919
30 LoZ
Capacitance
10.00
μ
F 9
μ
F
N/A
7.18 to 10.82
Set calibrator to standby, reconfigure leads, and program for amps output
31
A
DC Amps
10.00 A
10A
N/A
9.87 to
10.13
N/A
9.87 to
10.13
32
?
AC Amps
6.000 A
5.0A, 45 Hz
N/A
4.922 to
5.078
N/A
4.922 to
5.078
33
$
DC
μ
amps
600.0
μ
A 600
μ
ADC N/A
N/A
593.8 to
606.2
N/A
34
$
AC
μ
amps
600.0
μ
A 600
μ
AAC, 45 Hz
N/A
N/A
590.7 to
609.3
N/A
35
Open
input
N/A
N/A
0PEn
N/A
36
0.0
°
C
N/A
N/A
-1.0 to 1.0
N/A
37
y
Temperature
400
°
C N/A
N/A
395.0 to
405.0
N/A
38
0.5 V, 45 Hz
0.2 to 0.8,
AC
Annunciator
On
N/A
0.2 to 0.8,
AC
Annunciator
On
0.2 to 0.8,
AC
Annunciator
On
39
0.5 V, 0 Hz
0.2 to 0.8,
DC
Annunciator
On
N/A
0.2 to 0.8,
DC
Annunciator
On
0.2 to 0.8,
DC
Annunciator
On
40
x
500
V
[2]
,
500 Hz
489.7 to
510.3
N/A
489.7 to
510.3
489.7 to
510.3