Specifications

3

Specifications

General

Weight

280 g (10 oz)

Dimensions
(without holster)

2.8 cm

×

7.5 cm

×

16.6 cm

(1.1 in

×

3 in

×

6.6 in)

Battery

Standard 9 V battery (NEDA 1604, 6F22, or 006P)

Protection

Class III as defined in IEC 348, Safety Requirements for Electronic Apparatus

Environmental

Operating Temperature

0

o

C to +50

o

C (+32

o

F to +122

o

F)

Storage Temperature

40

o

C to +60

o

C (

40

o

F to +140

o

F)

Humidity

0

%

to 90

%

: 0

o

C to 35

o

C (+32

o

F to +95

o

F)

0

%

to 70

%

: 0

o

C to 50

o

C (+32

o

F to +122

o

F)

Electrical

Measurement Range

K-type:

200

o

C to +1370

o

C (

328

o

F to +2498

o

F)

J-type:

200

o

C to +760

o

C (

328

o

F to +1400

o

F)

Display Resolution

Low: +1.0

o

C or +1.0

o

F

High: +0.1

o

C or +0.2

o

F

Measurement Accuracy

K-type:

±

(0.1

%

of reading

+

0.7

o

C) [

±

(0.1

%

of reading

+

1.3

o

F)]

J-type:

±

(0.1

%

of reading

+

0.8

o

C) [

±

(0.1

%

of reading

+

1.4

o

F)]

Temperature
Coefficient

0.01

%

of reading

+

0.03

o

C (0.03

o

F) for [ambient temperatures from +18

o

C to +28

o

C

(+64

o

F to +82

o

F) ]

Maximum Differential Common Mode
Voltage

1 V (Maximum voltage difference between T1 and T2)

Temperature Scale

IPTS-68

Certification

P

,

$

Applicable Standards

NBS 125, IEC 584

Accuracy is specified for ambient temperatures between +18

o

C (+64

o

F) and +28

o

C (+82

o

F) for a period of 1 year. The above

specifications do not include thermocouple error.

Accuracy is unspecified when the input lead length is resonant with the interfering frequency.

Basic Maintenance

Basic maintenance required by the thermometer is discussed in this section.

Cleaning the Thermometer

Caution

To avoid damaging the thermometer, do not use aromatic hydrocarbons or

chlorinated solvents for cleaning. These solutions react with the plastics
used in the thermometer.

Do not allow the LCD to get wet.

Do not remove any of the lubricant from the input terminals.