Fluke 87 III User Manual
Page 54

80 Series III
Users Manual
46
Table 14. Current Function Specifications (continued)
Accuracy
1
Function
Range
Resolution
Model 83
2
Model 85
3, 4
Model 87
3, 4
Burden
Voltage
(typical)
µ
A
B
(45 Hz to 2 kHz)
µ
A
F
400.0
µ
A
4000
µ
A
400.0
µ
A
4000
µ
A
0.1
µ
A
1
µ
A
0.1
µ
A
1
µ
A
±
(1.2% + 2)
5
±
(1.2% + 2)
5
±
(0.4% + 4)
±
(0.4% + 2)
±
(1.0% + 2)
5
±
(1.0% + 2)
5
±
(0.2% + 4)
±
(0.2% + 2)
±
(1.0% + 2)
±
(1.0% + 2)
±
(0.2% + 4)
±
(0.2% + 2)
100
µ
V/
µ
A
100
µ
V/
µ
A
100
µ
V/
µ
A
100
µ
V/
µ
A
1.
See the first sentence in Table 11 for a complete explanation of accuracy.
2.
AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sinewave input.
3.
AC conversions for Models 85 and 87 are ac coupled, true rms responding, and valid from 5% to 100% of range.
4.
See note 3 in Table 11.
5.
Below a reading of 200 counts, add 10 counts.