Fluke 87 III User Manual
Page 52
80 Series III
Users Manual
44
Table 13. DC Voltage, Resistance, and Conductance Function Specifications
Accuracy
1
Function
Range
Resolution
Model 83
Model 85
Model 87
L
4.000 V
40.00 V
400.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
±
(0.1% + 1)
±
(0.1% + 1)
±
(0.1% + 1)
±
(0.1% + 1)
±
(0.08% + 1)
±
(0.08% + 1)
±
(0.08% + 1)
±
(0.08% + 1)
±
(0.05% + 1)
±
(0.05% + 1)
±
(0.05% + 1)
±
(0.05% + 1)
F
mV
400.0 mV
0.1 mV
±
(0.3% + 1)
±
(0.1% + 1)
±
(0.1% + 1)
e
nS
400.0
Ω
4.000 k
Ω
40.00 k
Ω
400.0 k
Ω
4.000 M
Ω
40.00 M
Ω
40.00 nS
0.1
Ω
0.001 k
Ω
0.01 k
Ω
0.1 k
Ω
0.001 M
Ω
0.01 M
Ω
0.01 nS
±
(0.4% + 2)
2
±
(0.4% + 1)
±
(0.4% + 1)
±
(0.7% + 1)
±
(0.7% + 1)
±
(1.0% + 3)
±
(1.0% + 10)
±
(0.2% + 2)
2
±
(0.2% + 1)
±
(0.2% + 1)
±
(0.6% + 1)
±
(0.6% + 1)
±
(1.0% + 3)
±
(1.0% + 10)
±
(0.2% + 2)
2
±
(0.2% + 1)
±
(0.2% + 1)
±
(0.6% + 1)
±
(0.6% + 1)
±
(1.0% + 3)
±
(1.0% + 10)
1.
See the first sentence in Table 11 for a complete explanation of accuracy.
2.
When using the REL
∆
function to compensate for offsets.