beautypg.com

Troubleshooting, Ce certification, Specifications – Fluke 561s User Manual

Page 15

background image

Troubleshooting

11

Troubleshooting

Symptom Problem

Action

--- (on display)

Target temperature is over or
under range

Select target within
specifications

B

Low battery

Replace battery

Blank display

Possible dead battery

Check and/or replace battery

Laser does not
work

1. Low or dead battery
2. Ambient temperature above
40

°

C (104

°

F)

1. Replace battery
2. Use in area with lower
ambient temperature

CE Certification

The Thermometer conforms to the following standards:

EN61326-1

EMC

Standard

EN61010-1

Safety

Standard

EN60825-1 Laser Standard

Certification testing was conducted using a frequency range of 80 to 1000 MHz with the
instrument in three orientations.

Specifications

Infrared

Measurement Range .................................................... -40

°

C to 550

°

C (-40

°

F to 1022

°

F)

Spectral Range............................................................. 8 to 14 microns
Accuracy ......................................................................

±

1 % or

±

1

°

C (2

°

F); <0

°

C (32

°

F),

±

1

°

C (2

°

F)

±

0.1

°

/1

°

(Assumes

ambient operating temperature of 23
to 25

°

C (73 to 77

°

F))

Repeatability.................................................................

±

0.5 % of reading or

±

1

°

C (2

°

F)

Display Resolution........................................................ 0.1

°

C (0.1

°

F)

Secondary Display Information..................................... Maximum, Minimum, Differential,

KTC

Response Time (95 %)................................................. 500 ms
Distance to Spot (D:S).................................................. 12:1
Emissivity Adjustment................................................... Three settings: low (0.3), medium

(0.7), high (0.95)

Contact Probe Input
Input Temperature Range ............................................ -40

°

C to 550

°

C (-40

°

F to 1022

°

F)

Input Accuracy...............................................................................Input accuracy

±

1.1

°

C (

±

2

°

F)

Display Resolution........................................................ 1

°

C (1

°

F)


Laser

Sighting ........................................................................ Single point laser
Power ........................................................................... Class 2 (II) operation; Output <1 mW,

wavelength 630 to 670 nm