There are connectors on the rear of the main display unit for 4 USB devices (flash drives, a QWERTY keyboard,
mouse, printer, barcode scanner), external speakers, an external HDMI monitor and connection to a network
(LAN) (serial number 4671 and higher). Units with serial number below 4671 have connectors for 2 USB
devices, an external VGA monitor, parallel printer, external speakers, a local area network (LAN) and an
RS223 serial port (not supported in this version of software).
Electrical supply input requirement: 100 – 240 Vac 47 – 63 Hz 250VA
A hospital grade grounded outlet is required.
Associated items and supplies
VA-100 Soft case for Verifit
VA-111 External speaker with folding stand and clips for microphone extension cables
VA-120 Barcode scanner for reading barcoded audiometric data on printouts from Audioscan analyzers
VA-131 Microphone extension cable (3') for use with Audioscan analyzers
VA-133 Microphone extension cable (10') for use with Audioscan analyzers
VA-201 NOAH
®
module allows a networked PC running NOAH to exchange data with Audioscan analyzers
RE201-25 Thermal printer paper for the internal printer in the Verifit
RE367-36 Probe microphone tubes for single patient use (36 per bag)
SAFETY WARNINGS and NOTICES
For purposes of IEC 60601-1, this product is Class I with Type BF applied part.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation
This Class A digital apparatus complies with Canadian ICES-003
WARNING: To avoid the risk of electric shock, this product must be connected only to a supply mains with
protective earth (ground).
WARNING: To avoid the risk of electrical shock, any line-powered peripheral equipment connected to this
product must comply with UL/IEC 60601-1 OR comply with UL or IEC and ISO safety standards for such
equipment AND a) be operated from an isolating transformer complying with UL/IEC 60601-1 OR b) be kept
at least 6 feet (1.8 m) from the patient.
WARNING: This equipment is not suitable for use in an oxygen-rich environment or in the presence of
flammable anesthetic mixtures with air or with oxygen or nitrous oxide.
WARNING: No modification of this equipment is permitted