Setting the pumping speed

45.0

MODE

➨ Press key MODE.

☞ Switching to vacuum control is only possible, if pumping

down is activated (after pressing key START).

☞ The actual pressure is stored as new set point, e. i. the con-

troller keeps the pressure constant.

☞ Press key MODE again to activate automatic mode.

Changing to mode vacuum control

Setting of speed:
Setting the value for the speed by using key Mode

▲ or ▼:

☞ To activate set mode: Press key shortly. The current speed

is displayed for one second.

☞ With a second tip within one second or continuous pressing:

☞ Set speed. The new set point is stored when the key is re-

leased.

☞ The maximum speed restricts the evaporation rate in case

of low boiling solvents and a carry over of substances.

☞ Setting the motor speed ”HI” (above 60 Hz) enables fast-

est pumping down to the best ultimate vacuum.