Manual determination of the set point p (e. g. boiling point)

➨ Setting the value for the speed by using key Mode ▲ or ▼

(factory-set: ”HI” = maximum motor speed).

☞ To activate set mode: Press key shortly. The current preset

speed is displayed for one second.

☞ With a second tip within one second or continuous pressing:

☞ Set speed. The new set point is stored when the key is re-

leased.

To start process control:
➨ Start: Press key START and immediately key p▼. Keep key

p

▼ pressed. Pumping down starts with preselected speed.

☞ Supervise process permanently. When sufficient evapora-

tion appears, release key p

▼.

☞ The actual pressure is stored as new set point.

☞ Vacuum control starts.

☞ If the automatic mode is activated additionally (key MODE)

the set point is adjusted automatically in case of change of
boiling point.

Setting of set point p (e. g. boiling point)

➨ Setting the set point by using the keys p▲ or p▼ (factory

set 100 mbar):

☞ To activate set mode: Press key shortly. “Set p“ appears.

☞ Short tips toggle p in steps of 1 mbar

☞ Continuous pressing activates ramp: Set-point changes with

increasing speed.

Adjusting the process pressure by pumping down:
➨ Press key p▼.

☞ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ With a second tip within one second or continuous pressing:

Pumping down with preset speed, the actual pressure is dis-
played and stored as new set point p when the key is re-
leased.

☞ The achieved pressure is stored as new set point and ad-

justed automatically.

Adjusting the process pressure upwards
➨ Press key p▲.

☞ When the key is pressed shortly, the current pressure is

displayed for one second.

☞ With a second tip within one second or continuous pressing:

The air admittance valve is opened if connected or pressure
increase due to system parameters, the current pressure is
displayed.

☞ The achieved pressure is stored as new set point when the

key is released and adjusted automatically (only in auto-
matic mode).

p

267

mbar

67

mbar

p

520

mbar

p

Adjusting of set point p during vacuum control

45.0

START

STOP

p