Aperture or lens wheel will not Dirt in or damage Service

turn or does so with difficulty to mechanism required

Visible corrosion on instrument Damage from excessive moisture Service required

Bent Arm Illuminator and Accessories Troubleshooting Guide

Trouble Area Possible Cause Corrective Action

Lamp burned out Replace the lamp with

ADC part #5211-4

No light

output Wrong lamp installed Replace the lamp with

ADC part #5211-4

Batteries are depleted

Replace with fresh “C” batteries

Accessories do not attach Accessories may have Obtain replacement

securely or operate properly reached end of useful life accessories

Instrument head will Thread/connection Service

not connect is damaged required

Lamp near end of Replace the lamp with

Dim light useful life ADC part #5211-4

output

Batteries are depleted

Replace with fresh “C” batteries

Visible corrosion on Damage from excessive Service

instrument moisture required

Environment

Transportation / Storage: -4°F to 120.2°F (-20°C to +49°C)

RH 95 %

500 hpa – 1060 hpa, Altitude

Environmental Operating Range: 50°F to 120.2°F (10°C to 49°C)

RH 95 % (Max)

500 hpa – 1060 hpa, Altitude

Maximum Operating Temperature:

95°F (35°C)

Symbols

Attention: Read Operating Manual for Cautions and Instructions for Use

.

Attention: This is NOT an AP or APG apparatus.

TYPE B. Indicates this is a product with Type B applied parts.