Instrument head Thread/connection Service

will is required

not connect damaged

Wrong specula Use only Welch Allyn brand

Specula will not used or compatible

attach properly

Specula mount is damaged Service required

Loose or misaligned Lens mount Service

magnifying lens is damaged required

Lens is dirty Clean with lint free cloth

Poor or obstructed view

through lens Lens is damaged Service required

Lamp near end Replace the lamp with
Dim light of useful life ADC part #5211-4

output

Batteries are Replace with fresh
depleted “C” batteries

Visible corrosion on Damage from excessive Service

instrument or accessories moisture required

Ophthalmoscope Troubleshooting Guide

Trouble Area Possible Cause Corrective Action

Aperture dial is Rotate Aperture dial

between positions

Lamp Burned Replace the lamp with

No light output out ADC part #5210-4

Wrong Lamp Replace the lamp with

Installed ADC part #5210-4

Batteries are depleted Replace with fresh “C” batteries

Instrument head

will Lamp not fully inserted Re-insert lamp

not connect

Thread/connection is damaged Service required

Spot is not The aperture dial is not centered Move aperture dial to

centered the full detent position

Lamp near end of useful life Replace the lamp with

Dim light ADC part #5210-4

output

Batteries are depleted Replace with fresh “C” batteries