6 technical specifications, 1 general, General – Metrohm Wall-Jet-Cell User Manual
Page 26
6.1 General
■■■■■■■■■■■■■■■■■■■■■■
20
■■■■■■■■
IC equipment Wall-Jet cell
6 Technical specifications
6.1
General
Structure
Flow-through cell with working, reference and auxiliary electrode.
Material
Cell body made of PEEK.
Cell volume
Dependent on the thickness of the spacer:
for 3 mm work-
ing electrode
and 50 µm
spacer
< 0.35 µL
for 2 mm work-
ing electrode
and 25 µm
spacer
< 0.1 µL
Maximum operat-
ing pressure
0.4 MPa (100 psi)
Cell recognition
Intelligent cell with automatic identification and monitoring.
Auxiliary electrode
Type
Installed
Material
Stainless steel
Reference elec-
trode
Type
Replaceable
Material
■
Palladium solid-state electrode
■
Ag/AgCl gel electrode