5 Operation and maintenance

IC equipment Wall-Jet cell

■■■■■■■■

13

5 Operation and maintenance

5.1

Operation

The measuring cell is operated together with the detector and the entire
IC system with the MagIC Net™ software.

Additional information on operation with MagIC Net™ can be found in
the "MagIC Net™ Tutorial" or in the MagIC Net™ online help.

5.2

Maintenance

The measuring cell must be taken out of the detector for the following
maintenance tasks.

CAUTION

The electrode cables may not be unplugged unless the measuring cell is
switched off in the software!

Taking the measuring cell out of the detector

1 Switch off the measuring cell in the software.

2 Disconnect all three electrode cables.

3 On the measuring cell holder, push the Press button and remove the

measuring cell from the holder.

5.2.1

Replacing the working electrode

Depending on the application, the measuring cell can be operated with a
variety of different working electrodes. The different working electrodes
are available as accessories (see Chapter 7, page 23).

Replacing the working electrode

Prerequisites:

The measuring cell is switched off.

The measuring cell has been removed from the detector.

The electrode cables are unplugged.