beautypg.com

2 replacing the reference electrode, Replacing the reference electrode – Metrohm Wall-Jet-Cell User Manual

Page 20

background image

5.2 Maintenance

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

IC equipment Wall-Jet cell

No tool is required for replacing the working electrode.

1 Unscrew the pressure screw (1-6) on the base part of the cell and

remove it.

2 Take out the working electrode.

3 Insert the new working electrode. The working electrode is shaped in

such a way that it can be inserted into the measuring cell in only one
position.

4 Slide the pressure screw over the working electrode and tighten it.

5.2.2

Replacing the reference electrode

Depending on the application, the measuring cell can be operated with a
variety of different reference electrodes. The different reference electrodes
are available as accessories (see Chapter 7, page 23).

NOTE

These instructions apply analogously to the Ag/AgCl reference elec-
trode.

Inserting the reference electrode

Prerequisites:

The measuring cell is switched off.

The measuring cell has been removed from the detector.

The electrode cables are unplugged.

No tool is required for replacing the reference electrode.

1 Slide the supplied sealing ring over the new reference electrode.