Barco MD-4221 User Manual
Page 12

12
RF Emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
D
Voltage fluctuations
IEC 61000-3-3
Complies
including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that
supplies buildings used for
domestic purposes
2. Guidance and manufacturer's declaration - electromagnetic immunity
The MD-4221 is intended for use in the electromagnetic environment specified below.
The customer or the user of the MD-4221 should assure that it is used in such an
environment.
Immunity test
IEC 60601 Test level
Compliance Electromagnetic
environment
-guidance
Electrostatic
discharge(ESD)
IEC61000-4-2
6kV Contact
8kV air
6kV Contact
8kV air
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%
Electrical fast
transient/burst
IEC 61000-4-4
2kV for power
supply lines
1kV for
input/output lines
2kV for power
supply lines
1kV for
input/output lines
Mains power
quality should be
that of a typical
commercial or
hospital
environment
Harmonic emissions
IEC 61000-3-2
D
Voltage fluctuations
IEC 61000-3-3
Complies