beautypg.com

Service log error messages appendix b – Anritsu 373XXA User Manual

Page 168

background image

SERVICE LOG ERROR MESSAGES

APPENDIX B

B-12

373XXA MM

Error Message

Description

Bad filename

The supplied filename was bad. The filename can have 8 characters maximum. No
extensions. The filename must start with and alpha type character (A thru Z). After
that the allowable characters are alpha, numeric (0 thru 9) and underscore (_).

Conflict with rotary knob

You should not be using the rotary knob and the GPIB at the same time.

Too many data points for exter-
nal source

A 6700B series external source can handle 501 data points.
A 68000 series external source can handle 999 data points.

Recalled setup corrupted
Hardware mismatch in recalled
setup
Software mismatch in recalled
setup

These are problems with the recalled setup.

Too many data points for Dis-
crete Fill

The maximum number of data points in discrete fill is 1601.

Not enough data points for Dis-
crete Fill

The minimum number or data points in discrete fill is 2.

Discrete Fill end frequency out of
range

The number of points for discrete fill puts the end frequency out of range.

Step is too large

When setting up a time domain harmonic sweep, cannot get 2 data points because
the start frequency is too high for the approximate stop frequency. In a group delay
display, the delay aperture percent of sweep is less than one step size.

Range too small

An attempt was made to set a distance or time span value too small. This can also be
done via inappropriate values for start and stop.

Start or stop out of range

An attempt was made to set a distance or time start or stop value out of range. This
can also be done via inappropriate values for center and span.

No bands defined

An attempt was made to turn on multiple source mode with no band definitions.

Out of frequencies for new band
definition
Source out of range by equation
External source out of range by
equation

The current set of multiple source mode bands use up all the frequency range of the
373XXA. Therefore, no more bands can be defined.

File is read only

An attempt was made to write to a write protected file.

File not found

An attempt was made to access a non-existent file.

Floppy drive not ready

An attempt was made to access the floppy drive with no floppy disk installed.

Floppy disk full
Hard disk full

An attempt was made to write to a floppy disk or the hard disk when no space was left
on the disk.

Table B-3. GPIB-Related Error Messages (5 of 8)