INOR MINIPAQ-L GB User Manual
Page 2

CALIBRATION
CONNECTIONS
OUTPUT LOAD DIAGRAM
It is recommended to check the calibration once a year. Use equipment
with an inaccuracy of at least 5 times better than the checked transmitter,
e.g. if you are checking a transmitter with an inaccuracy of maximum
0.1% use instrument with an inaccuracy of maximum 0.02%.
Use the MinIPAQ Soft Basic Calibration function in the Options menu, and
follow the instructions on the screen.
INPUT CONNECTIONS
DIMENSIONS
MOUNTING
+ -
RTD
T/C
RTD
RTD, 3-wire connection
Pt100, α =0.00385, Pt1000, α=0.00385
Pt100, α =0.003902, Pt100, α=0.003916
Ni100, Ni120, Cu10, Ni1000,
Pt(Spec), α=0.00385
Thermocouple
B, C, E, J, K, L, N, R, S, T, U
RTD
,
4-wire connection
Pt100, α=0.00385, Pt1000, α=0.00385
Pt100, α=0.003902, Pt100, α=0.003916
Ni100, Ni120, Cu10, Ni1000
Pt(Spec), α=0.00385
1
2
3
Output
PC
Connection
Input
4
21 22 23
Out
In
23
22
2
3
1
21
4
ORDERING INFORMATION
LIMITED WARRANTY
INOR Process AB, or any other affiliated company within the Inor Group (hereinafter
jointly referred to as ”Inor”), hereby warrants that the Product will be free from
defects in materials or workmanship for a period of five (5) years from the date of
delivery (”Limited Warranty”). This Limited Warranty is limited to repair or replace-
ment at Inor’s option and is effective only for the first end-user of the Product. Upon
receipt of a warranty claim, Inor shall respond within a reasonable time period as to
its decision concerning:
1 Whether Inor acknowledges its responsibility for any asserted defect in
materials or workmanship; and, if so,
2 the appropriate cause of action to be taken (i.e. whether a defective
product should be replaced or repaired by Inor).
This Limited Warranty applies only if the Product:
1 is installed according to the instructions furnished by Inor;
2 is connected to a proper power supply;
3 is not misused or abused; and
4 there is no evidence of tampering, mishandling, neglect, accidental damage,
modification or repair without the approval of Inor or damage done to
the Product by anyone other than Inor.
This Limited Warranty is provided by Inor and contains the only express war-
ranty provided.
INOR SPECIFICALLY DISCLAIMS ANY EXPRESS WARRANTY NOT
PROVIDED HEREIN AND ANY IMPLIED WARRANTY, GUARANTEE
OR REPRESENTATION AS TO SUITABILITY FOR ANY PARTICULAR
PURPOSE, PERFORMANCE, QUALITY AND ABSENCE OF ANY HIDDEN
DEFECTS, AND ANY REMEDY FOR BREACH OF CONTRACT, WHICH BUT
FOR THIS PROVISION, MIGHT ARISE BY IMPLICATION, OPERATION
OF LAW, CUSTOM OF TRADE OR COURSE OF DEALING, INCLUDING
IMPLIED WARRANTIES OF MER-CHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE. EXCEPT AS PROVIDED HEREIN,
INOR FURTHER DISCLAIMS ANY RESPONSIBILITY FOR LOSSES,
EXPENSES, INCONVENIENCES, SPECIAL, DIRECT, SECONDARY OR
CONSEQUENTIAL DAMAGES ARISING FROM OWNERSHIP OR USE
OF THE PRODUCT.
Products that are covered by the Limited Warranty will either be repaired or replaced
at the option of Inor. Customer pays freight to Inor, and Inor will pay the return
freight by post or other “normal” way of transport. If any other type of return freight
is requested, customer pays the whole return cost.
MinIPAQ-L
70MQL00003
Configuration
70CAL00001
ICON Configuration Kit (non-Ex)
70CFGUS001
+ -
RTD
T/C
RTD
Input connections
Output connections
35/1.38
17.5/0.69
68/2.68
63/2.48
98/3.86
45/1.
77
mm/inches
Mounting of MinIPAQ-L on rail DIN EN 50022, 35 mm.
1. Fix the upper part of the transmitter onto the rail.
2. Press the lower part of the transmitter against the rail.
3. To remove the transmitter, use a screwdriver and bend the
locking under the transmitter downwards.
R
LOAD
I
OUT
Test
1)
1)
Ri ≤ 10 Ω
Power Supply