beautypg.com

HT instruments HT157 User Manual

Page 9

background image

HT155 - HT157

EN - 7

4.2.

DESCRIPTION OF FUNCTION KEYS

Key Description

Pressing the ON key allows turning on and internally resetting
the instrument at any time

The OFF key allows turning off the instrument

The CANCEL key allows deleting the saved data from the
instrument’s memory

Activates/deactivates the display’s backlight. Backlight turns off
automatically after 15s

The ENTER key allows entering the internal programming
screens and confirming the selections made

The OUTPUT key allows stopping the measurements and
automatically saving the result in the memory

The START/PAUSE key allows activating or pausing a
measurement

,

,

,

The arrow keys allow navigating the programming menu,
increasing/decreasing the parameter values by one unit (press
and hold the keys for fast advance) and adjusting the display
contrast

The SETUP key allows rapidly accessing the programming
menu without passing from the main menu when a
measurement is in progress

4.3.

ANALOGUE OUTPUT DESCRIPTION

The instrument is provided with an AC voltage analogue output proportional to the input
signal measured in dB, which can be used to connect a 3.5mm output jack with the
characteristics shown in Fig. 3

Fig. 3: Description of jack for AC analogue output

The proportion between the output and the input signal depends on the measuring range
of the instrument and is shown in the following table:

Gain range

Output/input ratio

Measuring range [dB]

-10dB 0.316 50~140

0dB 1 40~130

10dB 3.16 30~120
20dB 10 20~110
30dB 31.6 10~100
40dB 100 0~90

The impedance of the AC output signal is 1k

. We recommend that the charge resistance

connected to the AC output is at least 100k

 and that the charge capacity is lower than

200pF. The ratio between the input and the output signal is 1mV / 0.1dB. The max output
voltage is 3.2V.

This manual is related to the following products: